Pages 67-68
The influence of sample volume applied to the Makler sperm counting chamber upon the measured concentration of latex beads: A multi-centre study
Authors: Melanie Walls, Emily Zuvela , Cheryl Ayres , Deborah Sherrin , Asma Chhotani , Liz Butler , Kelli Peirce , Jenny Krapez , Renae Parker , Cherise Mooy , Patrick Mohan , Sally Catt , Matthew Wiltshire, Hassan W Bakos, Mary Whyte, Phillip Matson
Number of views: 280
Objective
To undertake a multi-centre study to maximize the number of Makler chambers used.
Methods
A total of 15 laboratories participated with 31 Makler chambers. A suspension of latex beads was prepared to a concentration of 20 millions per milliliter, and 0.5 mL aliquots distributed to each participating laboratory. They measured the concentration on their Makler chamber(s) used for routine semen analysis by adding 3, 4, 5, 7 and 10 μL volumes of bead suspension to the chamber.
Results
There was no difference in within-chamber analysis of the bead concentration according to the volume of bead suspension applied within the range of 3–10 μL (F4,14=2.634, P=0.056). However, the between-chamber effects were significantly different (F30,124=4.937, P=0.000), and 24/31 (77.5%) chambers tested had an average bias>10% compared to the target bead concentration.
Conclusions
A volume of 3–10 μL added to Makler counting chambers does not influence the concentration measured of latex beads, but the between-chamber variability and positive bias seen would suggest that other sources of error are present which are yet to be identified.
Keywords
Volume;
Semen;
Makler chamber